Labotemp Natural Convention Drying Oven FOV-13 is a precision natural convection oven designed for controlled heating, drying and thermal processing in research laboratories, testing facilities and industrial quality control environments. With a 45 L chamber capacity and a temperature operating range of RT+10 to 300 °C, the unit supports routine analytical work and industrial oven drying tasks requiring a dependable scientific oven or laboratory drying oven.
The unit features a mirror stainless steel interior with a chemically resistant cold-rolled steel exterior, providing durability and easy cleaning during daily laboratory use. A natural convection heating system with vertical air circulation maintains uniform temperature throughout the chamber for reliable thermal performance. Built-in over-temperature protection, temperature-limiting alarm and reinforced door locking design ensure safe and stable operation, while the counter-rotating base plate design enables convenient sample handling.
Why choose Labotemp Natural Convection Drying Oven FOV-13?
✓ Stable Chamber Heating: Natural convection circulation maintains consistent temperature throughout the chamber.
✓ Advanced Microprocessor Control: PID-based regulation enables precise temperature management and repeatable results.
✓ Integrated Safety Protection: Over-temperature cutoff and alarm system ensure secure and reliable operation.
| No. | Specification | Value |
|---|---|---|
| 1 | Maximum Temperature | 300°C |
| 2 | Chamber Volume | 45 L |
| 3 | Power Rating | 1.0 kW |
| 4 | Temperature Distribution Accuracy | ±3.0% |
| 5 | Interior Material | Mirror stainless steel |
| 6 | Exterior Material | Cold-rolled steel with chemical-resistant coating |
| 7 | Insulation Material | Alumina silicate fiber |
| 8 | Heating Element | Finned stainless steel heating tube |
| 9 | Air Vent (Controller) | Ø28 mm × 1 (top) |
| 10 | Temperature Control Method | Dual PID digital tube display |
| 11 | Timer | 0–9999 minutes |
| 12 | Program Function | single-stage temperature control, 50 programs × 50 steps |
| 13 | Sensor | Pt100 |
| 14 | Safety Device | Anti-rise device |
| 15 | Internal Dimensions | 350×350×350mm |
| 16 | External Dimensions | 468×538×753mm |
| 17 | Package Dimensions | 592×618×880mm |
| 18 | Temperature Resolution | 0.1°C |
| 19 | Partition Load Capacity | 15 kg |
| 20 | Number of Layers | 6 / 2 |
| 21 | Partition Spacing | 40 mm |
| 22 | Power Supply | AC 220V / 50–60Hz 5.5A |
| 23 | Temperature Fluctuation | ±1.0°C |
| 24 | Gross Weight | 37kg |
| 25 | Net Weight | 43kg |
Used for precise drying, heating, sterilizing, and curing of samples in laboratories and research facilities thermal processing, temperature stability testing, and moisture removal, sample preparation, glassware drying, component aging tests